COMPARISON OF GEM PREMIER 5000 DEVICE AND RADIOMETER ABL 800 FLEX BLOOD GAS ANALYZERS
 
Yazarlar (2)
Doç. Dr. Muzaffer KATAR Tokat Gaziosmanpaşa Üniversitesi, Türkiye
Dr. Öğr. Üyesi Osman DEMİR Tokat Gaziosmanpaşa Üniversitesi, Türkiye
Makale Türü Açık Erişim Özgün Makale (Uluslararası alan indekslerindeki dergilerde yayınlanan tam makale)
Dergi Adı GEVHER NESIBE JOURNAL OF MEDICAL & HEALTH SCIENCES
Dergi ISSN 2717-7394
Dergi Tarandığı Indeksler Cite Factor, ESJI, Index Copernicus, Issuu Root Indexing
Makale Dili Türkçe Basım Tarihi 09-2023
Cilt / Sayı / Sayfa 8 / 0 / 639–648 DOI 10.5281/zenodo.8402027
Makale Linki https://gevhernesibedergisi.com/index.php/gnj/article/view/517/439
Özet
Objective: In this study, we aimed to compare the analytical performance of the Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA) blood gas analyzer with the ABL 800 Flex (Radiometer, Copenhagen, Denmark) blood gas analyzer used in our laboratory.Methods: 96 consecutive samples sent to Tokat Gaziosmanpaşa University Hospital Central Laboratory between 1-20 May 2023 were included in the study. Samples are taken into a 2 mL blood gas injector containing 70 IU of lithium heparin (Ayset Medical Products (Seyhan, Adana, Turkey) and analyzed with two devices consequetively. The parameters of pH, pCO2, pO2, ctHb, HCT, sO2, O2Hb, COHb, HHb, MetHb, Na, K+, Cl, Ca2+, Glucose, Lactate and HCO3 were evaluated. The significance test of the difference between the two spouses, Pearson correlation and Bland-Altman analysis were used. SPSS 19 (IBM SPSS Statistics 19, SPSS inc., An IBM Co., Somers, NY) was used.Results: In comparison of the two groups, clinically negligible, statistically significant difference was found between all parameters except pH, pO2, HCT, MetHb, Na, Cl and HCO3 (p< 0.001). Pearson correlation coefficients (CCC) of the parameters were in the range of 0.721-0.999. Bland-Altman values for pH, pCO2, MetHb and lactate were Bias: 0.013, SD: 0.031721, IIa:-0.04951, uIa: 0.07484, Bias-3.44286, SD: 3.378443, IIa:-10.0646, UIa: 3.178891, Bias:-0.1, SD: 0.676123, IIa:-1.4252, uIa: 1.225202, Bias: 0.269231, SD: 0.217503, IIa:-0.15708, ula: 0.695537 respectively.Conclusion: The GEM PREMIER 5000 was correlated with the Radiometer ABL800 Flex in terms of results and was acceptable …
Anahtar Kelimeler