Palatal perforation with McGrath Series 5 videolaryngoscope
   
Yazarlar (3)
Serkan Dogru
Prof. Dr. Serkan KARAMAN Tokat Gaziosmanpaşa Üniversitesi, Türkiye
Makale Türü Açık Erişim Özgün Makale (SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale)
Dergi Adı JOURNAL OF ANESTHESIA (Q4)
Dergi ISSN 0913-8668 Wos Dergi Scopus Dergi
Dergi Tarandığı Indeksler SCI-Expanded
Makale Dili İngilizce Basım Tarihi 04-2014
Cilt / Sayı / Sayfa 28 / 2 / 317–317 DOI 10.1007/s00540-013-1709-2
Makale Linki http://link.springer.com/10.1007/s00540-013-1709-2
Özet
To the Editor: A healthy pregnant patient (height 166 cm, weight 80 kg) who had an adequate preoperative fasting period was taken into the operating room to undergo an elective cesarean section. Induction of anesthesia was provided with thiopental sodium 5 mg/kg intravenously (IV) and rocuronium bromide 0.6 mg/kg IV. Direct laryngoscopy using a Macintosh 3 blade provided a Cormack–Lehane (CL) grade 3 glottic view. Tracheal intubation failed even when the head and neck were repositioned. A McGrath Series 5 videolaryngoscope (MGS-5; Aircraft Medical Limited, Edinburgh, UK) was used to see the glottis. However, no view was obtained from the first attempt. Nevertheless, intubation was successful during the second attempt, under a CL 2 view, with a styletted 7.5-mm endotracheal tube bent 60 upward just proximal to the cuff according to the angle of the blade. The tip of stylet did not protrude beyond the …
Anahtar Kelimeler
Palate | Soft | Laryngoscopy | Complications
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
Google Scholar 6
Web of Science 2
Palatal perforation with McGrath Series 5 videolaryngoscope

Paylaş