Highly selective potentiometric determination of nitrate ions using silver bisdiethyldithiocarbamate based membrane electrodes      
Yazarlar (2)
Prof. Dr. Ömer IŞILDAK Tokat Gaziosmanpaşa Üniversitesi, Türkiye
Doç. Dr. İlyas YILDIZ Iğdır Üniversitesi, Türkiye
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale
Dergi Adı Electrochimica Acta
Dergi ISSN 0013-4686 Wos Dergi Scopus Dergi
Dergi Tarandığı Indeksler SCI-Expanded
Dergi Grubu Q1
Makale Dili İngilizce
Basım Tarihi 01-2023
Cilt No 459
Sayı 1
Sayfalar 7 / 7
DOI Numarası 10.1016/j.electacta.2023.142587
Makale Linki http://dx.doi.org/10.1016/j.electacta.2023.142587
Özet
A new all-solid-state PVC membrane nitrate electrode was developed and nitrate ions were potentiometrically analyzed. Silver (I) bisdiethyldithiocarbamate ((C2H5)2NCS2)2Ag+) (Ag(dtc)2) compound was used as electroactive positive charged carrier (ionophore) in the plasticized PVC membrane structure of the developed electrode. The developed nitrate selective polymeric membrane electrode (nitrate-ISE) exhibited linear potentiometric behavior in the concentration range of nitrate ions from 1.0 × 10−1 to 1.0 × 10−5 mol L−1 and the detection limit was determined as 6.71 × 10−6 mol L−1. Although nitrate-ISE has good reproducibility, the pH working range is determined to be 4 – 9. The proposed nitrate-ISE has been successfully applied for the determination of nitrate ions in different water samples and fruit juices.
Anahtar Kelimeler
Nitrate ion-selective electrode | Potentiometry | PVC membrane