Özet
A novel poly(vinyl chloride; PVC) membrane‐based Cu(II)–selective sensor has been developed using synthesized 5,10,15,20‐tetrakis(4‐methoxycarbonylphenyl)porphyrin as an ionophore. The Cu(II)–selective sensor exhibits a Nernstian response for Cu(II) ion over a wide concentration range of 1.0 × 10−5 to 1.0 × 10−1 mol L−1 with the slope of 27.0 ± 2.5 mV/decade, and a lower limit of detection of 1.88 × 10−6 mol L−1. The novel sensor shows good repeatability and high selectivity for Cu(II) ions over other mono‐, bi‐ and trivalent cations. The sensor potential response remains unaffected by pH in the range of 4.4–10.6 and exhibits a fast response time of less than 10s. Finally the developed Cu(II)–selective sensor has been successfully used as an indicator electrode in the potentiometric titration of Cu2+ against ethylenediaminetetraacetic acid and also in the determination of Cu(II) ion …
Anahtar Kelimeler
copper(II) | porphyrins | potentiometry | PVC membrane | sensor