Silver(I)-selective PVC membrane potentiometric sensor based on 5,10,15,20-tetra(4-pyridyl)-21H,23H-porphine and potentiometric applications
Yazarlar (2)
Prof. Dr. Ömer IŞILDAK Tokat Gaziosmanpaşa Üniversitesi, Türkiye
Doç. Dr. Oğuz ÖZBEK Zonguldak Bülent Ecevit Üniversitesi, Türkiye
Makale Türü Açık Erişim Özgün Makale (SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale)
Dergi Adı Journal of Chemical Sciences (Q4)
Dergi ISSN 0974-3626 Wos Dergi Scopus Dergi
Dergi Tarandığı Indeksler SCI-Expanded
Makale Dili İngilizce Basım Tarihi 01-2020
Kabul Tarihi 09-10-2019 Yayınlanma Tarihi 25-01-2020
Cilt / Sayı / Sayfa 132 / 1 / 29–0 DOI 10.1007/s12039-019-1734-2
Makale Linki http://link.springer.com/10.1007/s12039-019-1734-2
Özet
In this study, all-solid-state contact PVC membrane silver(I)-selective sensor has been prepared and 5,10,15,20-tetra(4-pyridyl)-21H,23H-porphine used as ionophore. Optimum membrane composition; 3.5% ionophore, 59.6% Bis(2-ethylhexyl) sebacate (BEHS), 1.0% potassium tetrakis (p-chlorophenyl) borate (KTpClPB) and 35.9% polyvinyl chloride (PVC). The developed silver(I)-selective sensor exhibited a linear response (E = −51,57 (−log[Ag+]) + 1690,6 and R2 = 0,9986) in the varying concentration range of 1.0 × 10−1 to 1.0 × 10−5 mol L−1 silver ions. The limit of detection of the sensor was calculated as 1.9 × 10−6 mol L−1. The silver(I)-selective sensor exhibited good selectivity towards some alkali, alkali earth and heavy metal ions. The sensor can be used in the pH range from 4.0 to 10.0. The sensor has been shown to have good reusability and response time less than 8 s. The sensor was …
Anahtar Kelimeler
porphyrins | potentiometry | Sensors | silver | solid state