| Makale Türü | Özgün Makale (SCOPUS dergilerinde yayınlanan tam makale) | ||
| Dergi Adı | Bulgarian Chemical Communications | ||
| Dergi ISSN | 0861-9808 Scopus Dergi | ||
| Dergi Tarandığı Indeksler | Scopus | ||
| Makale Dili | İngilizce | Basım Tarihi | 12-2020 |
| Cilt / Sayı / Sayfa | 52 / 4 / 448–452 | DOI | 10.34049/bcc.52.4.5235 |
| Makale Linki | http://www.bcc.bas.bg/ | ||
| UAK Araştırma Alanları |
Analitik Kimya
|
||
| Özet |
| In this study, a novel poly (vinyl chloride) membrane bromide-selective sensor was developed. The best performance was obtained with a membrane composition of 2.0% ionophore, 33.0% poly (vinyl chloride)(PVC), 64.0% o-nitrophenyloctylether (o-NPOE) and 1.0% potassium tetrakis (p-chlorophenyl) borate (KTpClPB). The sensor exhibits linear potentiometric response towards bromide ion in the concentration range of 1.0× 10-1–1.0× 10-4 mol L-1 and has a detection limit of 2.2× 10-5 mol L-1. The bromide-selective sensor has a fast response time of 10 s and displays a highly selective response towards bromide compared to different anions. The sensor can be used in the pH range from 4.0 to 10.0. Finally, the bromide-selective sensor was successfully used for the determination of bromide in purified and commercial drinking water. |
| Anahtar Kelimeler |
| All-solid-state | Bromide determination | Ion-selective electrode | Potentiometry | PVC membrane |
| Atıf Sayıları | |
| Scopus | 11 |
| Google Scholar | 16 |
| Dergi Adı | Bulgarian Chemical Communications |
| Yayıncı | Bulgarska Akademiya na Naukite |
| Açık Erişim | Hayır |
| ISSN | 0861-9808 |
| E-ISSN | 2534-9899 |
| CiteScore | 0,7 |
| SJR | 0,148 |
| SNIP | 0,200 |