Özet
A novel potentiometric sensor based on ionophore (2-Hydroxymethyl-15-crown-5) for the determination of Fe3+ ions is developed. The Fe3+-selective membrane sensor demonstrates high sensitivity and short response time. The detection limit of the Fe3+-selective membrane sensor was about 1.2× 10− 6 mol L-1 and the response time was shorter than 8 s. The linear dynamic range of the Fe3+-selective membrane sensor was between 1.0× 10− 1 and 8.0× 10− 6 mol L-1 Fe3+ concentration. The Fe3+-selective membrane sensor exhibited good operational stability for at least two month keeping in dry conditions at 4–6◦ C. It had a reproducible and stable response during continuous standard deviation of 0.34%(n= 30).